Spezifikationen 100kHz DC

STANDARD SPECIFICATION FOR MODULE IN SIGNALOG 4000 SYSTEM CA 243
Typenumber: CA 243 Unit Function: DC-INSTRUMENTATION AMPLIFIER
Accuracy class % 0.05
Bandwidth (-3dB) kHz 100 (G = 20000); 140 (G = 2000)
Supply voltage V ± 15 (from dedicated powersupply)
Operating temperature °C 0 … 70
RANGE    
Range mV ± 0.5 / 1 / 2 / 5 / 10 / 20 / 50 / 100 / 200 / 500 / 1000 / 2000 / 5000
Gain Factor 20000 / 10000 / 5000 / 2000 / 1000 / 500 / 200 / 100 / 50 / 20 / 10 / 5 / 2
Digital fine-gain % 50 … 150 (in 4096 steps)
Gain accuracy % 0.05
Gain drift %/°C 0.0025
Linearity % 0.01
INPUT    
Impedance M Ohm 1
Capacitance pF 200
Offset voltage drift µV/°C 0,3 typical and 1 maximum
Offset current nA ± 25
Offset current drift pA/°C ± 15
Maximum common mode voltage V ± 15
Common Mode Rejection Ratio dB -90 ... -120 (dependent on range)
Noise RTI nV/Hz^0,5 5
Overload range   7x
CALIBRATION    
Calibration value mV 1
Calibration accuracy % 0.1
Input-zero suppression dB -90 ... -120 (dependent on range)
OPTION LE-FILTER    
Low-pass frequency (-3dB) Hz OFF / 0.8 / 1 / 1.25 / 1.6 / 2 / 2.5 / 3.15 / 5 / 8 / 10 / 12.5 / 16 / 20 / 25 / 31.5 / 50 / 80 / 100 / 125 / 160 / 200 / 250 / 315 / 500 / 800 / 1k/ 1k25 / 1k6 / 2k / 2k5 / 3k15 / 5k / 8k / 10k / 12k5 / 16k / 20k / 25k / 31k5 / 50k
Low-pass type   4-stage Besselfilter, -24dB/Oct
OUTPUT    
Full-scale voltage V ± 10
Maximum current mA 10
Protection   long-term short-circuit allowed
Impedance Ohm 0.5
Slew rate V/µs ± 10 (with 1kOhm load and G = 200 … 20000)
Overload recovery time µs 15 (5-times overload)
Maximum capacitive load nF 10
Maximum cable-length m 100 (100 pF/m)
Noise mVrms 2
GENERAL    
Standard Input-connector   7-pole MS-type (no 1/4-bridge + sensing simultaneously)
Alternative type   9-pole D-SUB
Advantage 9-pole D-SUB   1/4-bridge and 6-wire sensing simultaneously allowed
CURRENT OPTION 1    
Current output mA 0 … 20
Maximum voltage V 12
Impedance M Ohm 0.5
BRIDGE-SUPPLY OPTION    
Obtained strain-ranges mV/V ± 0.05 / 0.1 / 0.2 / 0.5 / 1 / 2 / 5 / 10 / 20 / 50 / 100 / 200 / 500 (bridge at 10 volts)
     
Voltage-mode ranges (DC) V 0 / 2 / 5 / 10
Voltage-mode max. current mA 100
Temperature drift %/°C 0.01
Noise voltage (until 10kHz) µV 100
     
Current-mode ranges (DC) mA 0 / 16.66 / 41.66 / 83.33 (optimized for 120 a. 350 Ohm brigdes)
Output - impedance k Ohm 200
     
Bridge-configurations   1/1-, 1/2-bridge (1/4-bridge only on D-type connector)
Half-bridge completion   120 Ohm, 0.02%, 2ppm
Straingauge impedances Ohm 60 … 3000
     
Accuracy (2-wire) % 0.1 (at bridge-connector not including cabling)
Accuracy (4-wire) % 0.05 (at bridge including cabling-resistance)
Maximum output-voltage V 11 (current-mode or 4-wire compensation)
Manual R-balance mV/V ± 0.6 or ± 6.0
     
Line-regulation 230V-bridge % 0.01 (or 400µV maximum)
TARE OPTION D253    
Digital tare   5 times mesuring range
Digital tare accuracy % 0.05
Resolution %, FS 0.02
Digital auto-tare time ms 10